    UV/VIS Spectrophotometer

Libra S80


Biochrom Libra S80 Double Beam Spectrophotometer





image

Features
 - Double beam UV/Vis spectrophotometer covering 190-1100nm
 - Variable bandwidth  (0.5nm, 1nm, 2nm, 4nm) and toluene in hexane ratio >2.0* for EP compliance
 - Compact design
 - Colour Touch screen display with easy to use software built in
 - USB connectivity for method and results storage plus connection to a PC
 - Resolution PC software included. Optional module for 21CFRpart 11 compliance
 (* at 1nm bandwidth)
 
Designed for busy multi-user laboratories
The Biochrom Libra S80 model is s robustly designed to deliver reliable, low maintenance UV/Visible spectrophotometry in environments with multiple users in a regulated environment such as method development. Users can quickly and easily walk up to the instrument, set up a sample, select the appropriate measurement protocols and collect the results.  Results and methods can be stored on board or the convenient USB port enables individual users to load personal methods and transport data. 21CFR Part 11 compliance via the optional Resolution software.
 
Easy to use
Colour touch screen user interface with menu driven software and simple intuitive icons is easy to both use and to learn. Results can be easily transported to a remote PC by memory stick for analysis later, or link to a PC directly with PVC software (included). Full PC control with Resolution software
 
Built in Software
Full suite of software applications provided as standard with real choice of data export methods. Trace Manager enables you to manipulate data stored on the instrument or USB memory stick. Multi level User log on facility allows supervisors to restrict access to software functionality allowing secure methods to be set up.



* Single Wavelength Measurement : Fixed wavelength, Abs/%T, and Concentration (factor)
* Wavelength Scanning : Sample overlays, automatic feature detection and post run data manipulation
* Concentration Standard Curve : Quantitation using calibration curves with multiple standards and a choice of curve fits
* Kinetics : Serial (or parallel measurements with optional cell changer) including sample overlays and post run data manipulation
* Life Science Methods : DNA, RNA & Oligo concentration and purity measurement with optional wavelength scanning. CyDyeTM DNA quantitation. Tm calculation. Protein concentration by direct UV measurement or Bradford, Biuret, Lowry or BCA colorimetric methods
* Equation Editor : Turn your instrument into a "custom application analyzer" by setting up stored methods including calculations based on measured sample data, for example: Absorbance difference and ratios.
 
Easy to Connect
Two USB ports are provided as standard. One is conveniently located by the user touchscreen to enable USB memory sticks (flash drives) to be be used to both store data and methods in secure and ASCII formats. Particularly useful for laboratories wishing to clone methods between multiple instruments. The other USB port is on the side of the instrument is used for PC communication, control, software update and print via computer (PVC). All stand-alone Biochrom Libra instruments offer users a choice in export and print functionality with selection parameters stored as part of any method.
 - Export to PC via USB cable (standard)
 - Built in printer (optional)
 - Export to a PC via Bluetooth (optional)
 - Full PC control with Resolution Software (included)
 
Resolution PC Software
Supplied with the Biochrom Libra S80 Resolution PC software offers users extra flexibility on data collection and manipulation. Resolution has the power to work in the way you want and includes modules for fixed wavelength, wavelength scanning, kinetics and quantitation measurements plus a validation routine for checking instrument performance. Data export options include Microsoft® Word and Excel® plus Adobe® PDF formats. Optional upgrade modules are also available including Resolution Life for Nucleic Acid and protein measurement and Resolution CFR for 21 CFR part 11 compliance.
 
Superb Optical Design
High energy throughput double beam optical system with variable bandwidth (0.5nm, 1nm, 2nm, 4nm) gives users the flexibility and performance to measure demanding sample types whilst ensuring EP compliance.
 
Cell Holders
The Biochrom Libra S80 is supplied with 10mm pathlength cell holders in sample and reference positions.



Specifications
Type : UV/Vis
Wavelength Range : 190 - 1100nm
Wavelength Accuracy (+/-) : ± 0.3nm across wavelength range (at 1nm bandwidth)
Wavelength Reproducibility (+/-) : ± 0.1nm (at 1nm bandwidth)
Lamp Source : Deuterium/Tungsten
Beam : Double Beam
Optical System : Czerny-Turner Monochromator
Bandwidth : Variable (0.5nm, 1nm, 2nm, 4nm)
Detector : Dual Solid State Silicon photodiode
Photometric Range : -4.000A to 4.000A
Stray Light : <0.025%T at 220nm (NaI), <0.025%T at 340nm (NaNO2), <1%T at 198nm & 200nm (KCl) (at 1nm bandwidth)
Zero Stability : ± 0.0003A/hr at 500nm in precision mode (at 1nm bandwidth)
Baseline Flatness : ± 0.001A 220 to 900nm (at 1nm bandwidth)
Measurement Type : Fixed Wavelength, Abs/%T, Abs ratio, Abs difference, concentration (standard curve), concentration (factor), multiwavelength, kinetics, Life Science Methods, user log-on, Equation editor
Cell Compatibility : Cuvette, Microcell*, Variable Pathlength Cell*
Other Sample Containers :  Test Tube*
Cell Holder : Dual (sample and reference), 8 cell changer*, 5 position heated cell changer*, variable pathlength cell holder*
Accessory Options : Film holder*
Temperature Control : Temperature controlled cell changer*, Water Jacket*
Method Storage : Up to 90 on-board, unlimited via USB
Regulatory Compliance : European Pharmacopoeia, IQ/OQ*, 21CFR part 11*
Printer : Built in Thermal*, External via Export*
Data Output : Store, Print*, USB Drive*, Export to PC, PC Control*
Display : Colour Touchscreen
Export/ Data Storage : Internal Memory, USB Drive*, Bluetooth*, USB cable to PC
Data Interface : USB
Languages : English, French, German, Spanish
Power Requirement : 85-264 V (150 VA)
Weight : 18Kg (39.7lb)
Dimensions (WxDxH) : 535 x 440 x 335mm (21.3 x 18.1 x 13.2 inches)
Certification : CE
Noise : <0.002A peak to peak, 0.0005A RMS near 2A at 500nm. <0.00005A RMS at 0A at 500nm over 20 measurements (at 1nm bandwidth)
Photometric Accuracy : ± 0.002A at 0.5A, ± 0.004A at 1A, ± 0.006A at 2A at 440nm, 456nm, 546.1nm, 590nm, 635nm ± 0.005A using 60mg/L K2Cr2O7
Photometric Reproducibility : ± 0.002A at 1A (at 1nm bandwidth)
Scan Speed : >2400 nm/min
Spectral Scanning : Yes (up to 1100nm)



Ordering Information
Biochrom Libra S80 UV/Vis Spectrophotometer    80-7000-31
Biochrom Libra S80 UV/Vis Spectrophotometer (with built in printer)    80-7000-32
Biochrom Libra S80 UV/Vis Spectrophotometer (with Bluetooth®)    80-7000-33
Biochrom Libra S80 UV/Vis Spectrophotometer(with built in printer and Bluetooth®)    80-7000-34



Spare Parts
80-7000-50 Spare Deuterium Lamp
80-7000-51  Spare Tungsten Lamp
80-3004-07 Spare Printer paper for Built in Printer (20 rolls)